Custom service
Custom Membrane Protein QC
Proof that a membrane protein preparation is what it claims to be.
A membrane protein batch can pass every standard protein QC check and still be unusable. SDS-PAGE says the protein is there; SEC says it is mostly monomeric; and neither says whether the conformation that binds ligand survived the extraction. Membrane protein QC therefore has to ask a different set of questions, and answer them with methods designed for the question.
The panel runs on four levels. Identity: intact mass and peptide mapping confirm the construct, its modifications and its processing. Physical state: SEC-MALS gives mass and oligomeric state in the native detergent or nanodisc; DLS catches aggregate that SEC misses. Stability: nanoDSF or CPM gives the unfolding temperature, and a freeze–thaw or time-course stability study shows how the protein behaves in storage. Function: ligand binding, enzymatic activity, transport or signalling — the readout that decides whether the batch is fit for its purpose.
Function is the part most often skipped and most often missed. A batch report without a functional measurement describes a protein's appearance, not its usability. Where no functional assay exists for the target, one is built or the report states plainly what is unverified.
Membrane protein QC is used before an expensive downstream step — immunisation, structure determination, assay transfer — and when a batch from another source must be verified before it is trusted.
It is not a fix for a failed preparation. QC finds problems; it does not solve them, and a failed QC report is the cheaper outcome when it arrives before the next stage rather than after.
The QC report states every method, its acceptance criterion and the measured result. Where a criterion is not met, the failure is reported rather than re-averaged. Raw data — chromatograms, spectra, DSF traces — accompany the report.
Service Highlights
- Identity by mass spectrometry. Intact mass and peptide mapping confirm construct, tag cleavage, disulfide state and key modifications.
- Mass and oligomeric state. SEC-MALS in the delivery format, not in a different detergent, so the reported state matches the product.
- Two stability methods. NanoDSF and CPM run in parallel to catch proteins that one dye misses.
- Storage stability. Freeze–thaw cycles and time-course incubation at the storage condition, with monomer content tracked.
- Functional readout required. Ligand binding, activity or transport measured on the batch; a report without function states what was not verified.
- Format-aware methods. Nanodisc and liposome QC uses methods that work in those formats rather than forcing the protein back into detergent.
How the work runs
QC plan
The batch's intended use defines the panel — which methods, which acceptance criteria and which functional readout.
Identity
Intact mass and peptide mapping confirm construct and modifications.
Physical characterisation
SEC-MALS and DLS for mass, oligomeric state and aggregation.
Stability
Thermal transition by nanoDSF and CPM, plus storage stability as required.
Function
The agreed functional measurement run on the batch, compared with reference material where it exists.
Report
A QC report with every measurement, the acceptance criteria and a statement of what was and was not verified.
Specifications & timeline
- Identity
- Intact mass LC-MS, peptide mapping
- Physical
- SEC-MALS, analytical SEC, DLS, SDS-PAGE
- Stability
- nanoDSF, CPM assay, freeze–thaw and storage time-course
- Function
- Ligand binding, enzymatic activity, transport, signalling — per target
- Timeline
- 1–2 weeks per batch
Related Products
- Recombinant ProteinsGrowth factors, cytokines and antigens produced across E. coli, yeast, insect and mammalian hosts — over 10,000 items on the shelf and ready to move.
- Native ProteinsIsolated directly from natural sources, preserving native post-translational modifications and conformational epitopes that recombinant systems can miss.
- Monoclonal AntibodiesRecombinant and hybridoma clones characterised for Western blot, IHC, immunofluorescence, ELISA and flow cytometry — application data included.
Next step
Tell us the target, we will tell you what is feasible
Send the sequence, assay format and sensitivity you need. You will get a technical assessment — recommended platform, timeline and the risks worth knowing before you commit budget.