Custom service
Mass Standard Protein Production
Protein whose concentration you can defend, because it was assigned and verified.
A mass standard is a protein lot whose concentration is known well enough to be trusted as a reference. Most laboratory protein preparations are not: they are assigned by A280, which inherits the error of the extinction coefficient, or by BCA, which inherits the response of a calibration protein that is not the analyte. The result is a reagent whose true concentration may be ten or twenty percent away from the value on the label, and every assay built on it inherits that error.
Production of a standard starts with the protein, not the metrology. It is expressed in a host chosen for clean glycosylation or none, purified to above 98% monomer, and verified for identity by intact mass and peptide mapping. Only then is concentration assigned: amino acid analysis for absolute quantitation, supported by a gravimetric preparation and, where the protein allows, by orthogonal methods such as quantitative NMR or isotope-dilution mass spectrometry.
Stability is part of the product. A standard is aliquoted, lyophilised or stored at −80 °C, and held with a stability protocol that measures monomer content and concentration at defined intervals. What is delivered is a lot certificate with the assigned value, its uncertainty, and the storage conditions that keep it true.
Mass standards are needed wherever a concentration value is load-bearing: assay calibrators, instrument standards for mass spectrometry and SEC, inter-laboratory comparison samples, and reference materials for regulated work.
They are not needed for routine reagent production, where A280 assignment carries an acceptable error, and the metrology cost is difficult to justify for a protein used once in a pilot experiment.
The lot certificate states the assigned concentration, its uncertainty, the assignment method, purity by SEC-HPLC, identity by mass spectrometry, endotoxin where measured, and the storage condition with its stability protocol. Holding samples are retained for lot verification on request.
Service Highlights
- Assigned by amino acid analysis. Acid hydrolysis followed by HPLC or UPLC quantitation of stable residues, with uncertainty stated rather than implied.
- Identity verified. Intact mass by LC-MS and peptide mapping confirm the sequence before any value is assigned.
- Purity above 98% monomer. SEC-HPLC with the chromatogram in the certificate; aggregates are quantified, not hidden.
- Endotoxin controlled. Below 0.1 EU/mg where the standard is used in cell-based or biological assays.
- Aliquoted and stability-tested. Lyophilised or frozen aliquots with a documented stability protocol and holding samples.
- Certificate of analysis. Assigned value, uncertainty, method, buffer composition and storage condition for every lot.
How the work runs
Target specification
Protein, required purity, matrix and intended use as a calibrator are agreed, and the acceptance criteria are written before production.
Expression and purification
Host selected for clean modification; purification to above 98% monomer with aggregate removal documented.
Identity and purity verification
Intact mass, peptide mapping, SDS-PAGE and SEC-HPLC establish identity and purity before metrology.
Value assignment
Amino acid analysis as the primary method, with gravimetric preparation and orthogonal confirmation where the protein permits.
Aliquoting and storage
Sterile aliquoting into the required format, lyophilisation or freezing, with holding samples retained.
Stability protocol
Monomer content and concentration measured at defined intervals; the stability plan and first data point are included in the lot file.
Specifications & timeline
- Purity
- >98% monomer by SEC-HPLC
- Assignment
- Amino acid analysis; uncertainty typically ±2–5%
- Identity
- Intact mass and peptide mapping by LC-MS
- Endotoxin
- Below 0.1 EU/mg on request
- Format
- Liquid aliquots or lyophilised vials
- Timeline
- 4–8 weeks depending on protein and metrology depth
Related Products
- Recombinant ProteinsGrowth factors, cytokines and antigens produced across E. coli, yeast, insect and mammalian hosts — over 10,000 items on the shelf and ready to move.
- Native ProteinsIsolated directly from natural sources, preserving native post-translational modifications and conformational epitopes that recombinant systems can miss.
- Monoclonal AntibodiesRecombinant and hybridoma clones characterised for Western blot, IHC, immunofluorescence, ELISA and flow cytometry — application data included.
Next step
Tell us the target, we will tell you what is feasible
Send the sequence, assay format and sensitivity you need. You will get a technical assessment — recommended platform, timeline and the risks worth knowing before you commit budget.